Flow Cytometry
Flow Cytometry shows the individual samples your Bactosense instruments measure: when each sample was taken, how many cells it counted, and the plots the counts come from. Open it from the left navigation.
Everyone with access to the workspace can read this screen and download from it. Changing a gate is for the Manager and Admin roles — see How gating works.
One series selected over 30 days. The list on the left, the selected sample on the right:
its details, four result tiles and three plots, with Analyze sample above them for a Manager.
What the numbers mean
Section titled “What the numbers mean”A flow cytometer stains the cells in a water sample and counts them one by one as they pass a laser. Each particle that passes is an event, with a green fluorescence (FL1), a red fluorescence (FL2) and light scatter (SSC, side scatter; FSC, forward scatter). The gate is the region of the FL1 × FL2 plot whose events count as cells; everything outside it is noise, debris or background.
| Tile | Meaning |
|---|---|
| TCC | Total cell count, in cells/mL. Shown for samples of type TCC-D, where all cells are stained. |
| ICC | Intact cell count, in cells/mL. Shown instead of TCC for samples of type ICC-A, which count only cells with an intact membrane. |
| HNA | High nucleic acid cells: counted cells to the right of the HNA split on FL1. Often the more active, faster-growing bacteria. |
| LNA | Low nucleic acid cells: counted cells to the left of the split. |
| HNAP | The HNA share of the count, as a percentage. |
| ACC | Additional cell count: a second, optional region of the plot for a population you want counted separately. Only shown when the gate defines one. |
For a stable water source the pattern of the plots — its fingerprint — changes slowly. A shift in HNAP or a new cloud of events is often an earlier sign of change than the total count.
The filter bar
Section titled “The filter bar”| Control | What it does |
|---|---|
| Series | Chooses which series to show. Switch between View by location and View by tag, search, and tick one or more. With nothing ticked, the list shows every flow-cytometry sample in the workspace. Series without samples in the period are hidden; Show series without data in range shows them. |
| Period | The date range, Last 30 days by default. Pick Last day, Last 7 days, Last 30 days, Last 90 days or Last 365 days, or select two days in the calendar and select Apply. Days start at midnight in the workspace’s time zone, named at the bottom. |
| Play () | Play through samples: steps through the selected sample’s series, one sample a second, toward the newest. See Review samples. |
The selected series, the period and Show series without data in range are remembered for the browser tab. The open sample is part of the page’s web address, so a link to the page opens that sample. The list refreshes itself every minute; Refresh in the page’s menu (⋮) does it at once.
Export results at the top right downloads the results as an Excel file — see Export results and FCS files.
Samples
Section titled “Samples”Each card shows the device, the series’ tag, the location, the date and time, and the primary count (TCC or ICC) in cells/mL, newest first. The list holds 100 samples per page; the pager under it shows the page number, and the download icon () beside it downloads the FCS files of the page.
Marks on a card:
| Mark | Meaning |
|---|---|
| Orange triangle | The instrument reported a warning that may affect accuracy. The details are at the bottom of the sample. |
| (gray) | The sample is queued to be recomputed with a newer gate. The number shown is still the old one. |
| (orange, with a mark) | The recompute failed (Recomputing failed…). The sample keeps its previous numbers. |
The selected sample
Section titled “The selected sample”The heading names the location and device, with the tag. Beside it are the same warning and recompute marks, the Analyze sample button (Manager and Admin), and the sample’s menu (⋮).
Details
| Field | Meaning |
|---|---|
| Sample name | The name the instrument gave the sample. |
| Datetime | When it was measured. The arrows beside it step to the previous and next sample of the same series. |
| Type | TCC-D (total cell count) or ICC-A (intact cell count). |
| Gate | The gate that counted this sample — see below. |
| Cartridge Serial, Fill Serial | The reagent cartridge and fill the instrument used. |
| Volume | The volume analyzed, in µL. |
| Events | The number of events the instrument recorded. |
Results: the tiles described above.
Plots
- FL1 × FL2 — the density plot of every event, with the gate outline, the ACC region if there is one, and the HNA split as a vertical line.
- A cross-scatter plot of the counted events: FL1 × SSC, FL1 × FSC or SSC × FSC, depending on which scatter channels the instrument records.
- FL1 histogram — the counted events along FL1, colored LNA and HNA either side of the split.
A sample the server has not processed yet shows No plots for this sample.
Which gate counted it
Section titled “Which gate counted it”The Gate field names the gate behind the numbers. The first of these that applies wins:
| Label | Meaning |
|---|---|
| Hand-gated, with This sample only | A gate was drawn for this sample alone. It no longer follows its series. |
| gate name vn, for example Intake TCC v2 | A gate from the workspace’s gate library, at the version that counted this sample. |
| Device gate | The gating the instrument sent with the sample. The series follows no library gate. |
| Default gate | The platform’s fallback, used when nothing else defines a region. |
How gating works explains the order.
Calculation history
Section titled “Calculation history”A sample that has been recomputed shows Calculation history under its plots, with the number of calculations. Open it to see every calculation, newest first, marked Current at the top. Each row gives when it ran, the gate it ran under (Hand-gated if the sample had its own gate), its results, and how much each result differs from the row below it. Point at a gate name to see that gate drawn on this sample’s FL1 × FL2 plot and histogram; click it to keep it open.
A sample calculated three times: first with the device gate, then twice with the library
gate Intake TCC v2. Red and green figures are the change from the row below.
A sample that was never recomputed has no calculation history section.
Warnings
Section titled “Warnings”When the instrument raised warnings, an orange box at the bottom of the sample lists each one by code and description: This sample contains warnings that may affect accuracy. The result is still shown; read it with care.
The sample menu
Section titled “The sample menu”The menu (⋮) beside the sample holds:
| Item | What it does | Who |
|---|---|---|
| Download sample file | Downloads this sample’s FCS file. | Everyone |
| Use the series’ gate | Only on a hand-gated sample: drops its own gate and recomputes it with its series’ gate. See Hand-gate a sample. | Manager, Admin |
| Analyze sample (desktop) | On a narrow screen the gate editor is not available; the item is shown grayed out. | Manager, Admin |
Use the series’ gate is there only because this sample is hand-gated, and only for a Manager or Admin.
On a phone
Section titled “On a phone”On a narrow screen the list and the sample take turns: tap a sample to open it full screen, with Previous sample and Next sample links, and the back arrow to return to the list. Export results moves into the page’s ⋮ menu. The Regate requests drawer is in the user menu (your initials).